PP201
The PP201 is a 4-quadrant power potentiostat designed to apply and sink high currents up to ±20 A at a voltage range of ±10V. The total power dissipation of the PP201 is 200W.
The PP-Series potentiostats are controlled by an EPC42, a plug-in module for the ZENNIUM-series Electrochemical Workstations. Due to the built-in microprocessor, each EPC42 can address up to four PPSeries potentiostats. Up to four EPC42 cards can be installed in an ZENNIUM-series workstation. In total, up to 16 PPs can be controlled by one ZENNIUM. Each potentiostat will hold the control parameters from one access to the next one, so that no potential or current disturbances can occur while scanning the potentiostats. If series measurements are performed with more than one PP-Series potentiostat, spectra are taken from all modules in a definable order.
The PP-Series is embedded completely in the ZENNIUM environment. Thus, all acquisition and analysis techniques that run on the ZENNIUM workstation can be applied with the power potentiostats as well. The installation of one or more PP-Series potentiostat will upgrade your ZENNIUM to an even more versatile, high-current electrochemical workstation.
The PP-Series potentiostats can also be controlled by a Windows®-PC. In this case they provide methods, summarized in the table below. They also work as a LabView® Virtual Instrument under the LabView® software. To implement the PP-Series potentiostats into existing test environment, a supporting DLL is available on demand.
You can control the PP-Series potentiostats in a mixed mode with an ZENNIUM and a PC in parallel. Both devices can be connected and disconnected during operation.
Supported methods with ZENNIUM-series |
Software module |
Impedance spectroscopy |
IM |
Simulation & model fitting |
SIM |
Cyclic voltammetry |
CV |
Polarisation curves |
I/E |
Multicell multitasking voltammetry |
CV |
Arbitrary current/potential/time meas. |
PVI |
Capacity/potential measurements |
C/E |
Automatic series measurements |
AS |
Supported methods with PC |
Test Sampling |
U vs. time, I vs. time |
Current Potential Curves (U/I) |
Cyclic Voltammetry, CV at OCP |
Charging/Discharging, Battery Cycling |
LabView® Virtual Instrument |
DLL support available |
PP201 |
|
Operating modes |
pot/gal/oc |
Potential range |
±10 V |
Potential accuracy |
±0.1% / ±1 mV |
Current range |
0 A … ±20 A |
Current accuracy |
±0.25 % / ±1 mA |
Output power |
200 W |
Frequency range |
10µHz - 200kHz |
Impedance range |
1µ Ω - 1 kΩ |
Ambient temperature |
0oC … 30oC |
Dimensions |
364 x 160 x 376 mm |
Weight |
10kg |
System requirements |
ZENNIUM workstation+EPC42 or PC |
Brochure PP2x1
Download the latest brochure about ZAHNER PP2x1...
b_pp.pdf (PDF)
Manual PP-Series / EL / XPOT
Download the latest manual about ZAHNER PP2x1/XPot...
m_ppxpot.pdf (PDF)
PP-Inspector V6.4 (PC Software)
Download the latest version of PP-Inspector software...
PPInspector_64.exe (Setup)
Manual PP-Inspector
Download the latest manual about PP-Inspector software...
m_ppinspector.pdf (PDF)