Genauigkeit des Power Potentiostate

XPOT2 – Voltage Ranges
Voltage range indexRange Standalone [V] minRange Standalone [V] maxRange Thales [V] minRange Thales [V] maxTolerance
0-5.05.0-4.04.0±500 µV ±0.1% of reading
1-12.512.5-10.010.0±1250 µV ±0.1% of reading
XPOT2 – Current Ranges
Shunt indexRange Standalone [A] minRange Standalone [A] maxRange Thales [A] minRange Thales [A] maxTolerance
0-500m500m-500m500m±250 µA ±0.1% of reading
1-160m160m-160m160m±85 µA ±0.1% of reading
2-40m40m-40m40m±20 µA ±0.1% of reading
3-5m5m-4m4m±2.5 µA ±0.1% of reading
4-500µ500µ-400µ400µ±250 nA ±0.1% of reading
5-50µ50µ-40µ40µ±25 nA ±0.1% of reading
6-5µ5µ-4µ4µ±2.5 nA ±0.1% of reading
7-500n500n-400n400n±250 pA ±0.1% of reading
PP212 20 V 10 A – Voltage Ranges
Voltage range indexRange Standalone [V] minRange Standalone [V] maxRange Thales [V] minRange Thales [V] maxTolerance
0-1010-88±1000 µV ±0.1% of reading
1-2020-2020±2500 µV ±0.1% of reading
PP212 20 V 10 A – Current Ranges
Shunt indexRange Standalone [A] minRange Standalone [A] maxRange Thales [A] minRange Thales [A] maxTolerance
0-1010-1010±100 mA ±0.2% of reading
1-1010-1010±10 mA ±0.2% of reading
2-22-22±2 mA ±0.2% of reading
3-200m200m-200m200m±200 µA ±0.2% of reading
PP222 10 V 20 A – Voltage Ranges
Voltage range indexRange Standalone [V] minRange Standalone [V] maxRange Thales [V] minRange Thales [V] maxTolerance
0-55-44±500 µV ±0.1% of reading
1-1010-1010±1250 µV ±0.1% of reading
PP222 10 V 20 A – Current Ranges
Shunt indexRange Standalone [A] minRange Standalone [A] maxRange Thales [A] minRange Thales [A] maxTolerance
0-2020-2020±200 mA ±0.2% of reading
1-2020-2020±20 mA ±0.2% of reading
2-22-22±2 mA ±0.2% of reading
3-200m200m-200m200m±200 µA ±0.2% of reading
PP242 5 V 40 A – Voltage Ranges
Voltage range indexRange Standalone [V] minRange Standalone [V] maxRange Thales [V] minRange Thales [V] maxTolerance
0-55-44±500 µV ±0.1% of reading
1-55-55±1250 µV ±0.1% of reading
PP242 5 V 40 A – Current Ranges
Shunt indexRange Standalone [A] minRange Standalone [A] maxRange Thales [A] minRange Thales [A] maxTolerance
0-4040-4040±400 mA ±0.2% of reading
1-4040-4040±40 mA ±0.2% of reading
2-22-22±2 mA ±0.2% of reading
3-200m200m-200m200m±200 µA ±0.2% of reading

Fordern Sie ein Angebot an