Accuracy of Power Potentiostats
XPOT2 – Voltage Ranges Voltage range index Range Standalone [V] min Range Standalone [V] max Range Thales [V] min Range Thales [V] max Tolerance 0 -5.0 5.0 -4.0 4.0 ±500 µV ±0.1% of reading 1 -12.5 12.5 -10.0 10.0 ±1250 µV ±0.1% of reading
XPOT2 – Current Ranges Shunt index Range Standalone [A] min Range Standalone [A] max Range Thales [A] min Range Thales [A] max Tolerance 0 -500m 500m -500m 500m ±250 µA ±0.1% of reading 1 -160m 160m -160m 160m ±85 µA ±0.1% of reading 2 -40m 40m -40m 40m ±20 µA ±0.1% of reading 3 -5m 5m -4m 4m ±2.5 µA ±0.1% of reading 4 -500µ 500µ -400µ 400µ ±250 nA ±0.1% of reading 5 -50µ 50µ -40µ 40µ ±25 nA ±0.1% of reading 6 -5µ 5µ -4µ 4µ ±2.5 nA ±0.1% of reading 7 -500n 500n -400n 400n ±250 pA ±0.1% of reading
PP212 – Voltage Ranges Voltage range index Range Standalone [V] min Range Standalone [V] max Range Thales [V] min Range Thales [V] max Tolerance 0 -10 10 -8 8 ±1000 µV ±0.1% of reading 1 -20 20 -20 20 ±2500 µV ±0.1% of reading
PP212 – Current Ranges Shunt index Range Standalone [A] min Range Standalone [A] max Range Thales [A] min Range Thales [A] max Tolerance 0 -10 10 -10 10 ±100 mA ±0.2% of reading 1 -10 10 -10 10 ±10 mA ±0.2% of reading 2 -2 2 -2 2 ±2 mA ±0.2% of reading 3 -200m 200m -200m 200m ±200 µA ±0.2% of reading
PP222 – Voltage Ranges Voltage range index Range Standalone [V] min Range Standalone [V] max Range Thales [V] min Range Thales [V] max Tolerance 0 -5 5 -4 4 ±500 µV ±0.1% of reading 1 -10 10 -10 10 ±1250 µV ±0.1% of reading
PP222 – Current Ranges Shunt index Range Standalone [A] min Range Standalone [A] max Range Thales [A] min Range Thales [A] max Tolerance 0 -20 20 -20 20 ±200 mA ±0.2% of reading 1 -20 20 -20 20 ±20 mA ±0.2% of reading 2 -2 2 -2 2 ±2 mA ±0.2% of reading 3 -200m 200m -200m 200m ±200 µA ±0.2% of reading
PP242 – Voltage Ranges Voltage range index Range Standalone [V] min Range Standalone [V] max Range Thales [V] min Range Thales [V] max Tolerance 0 -5 5 -4 4 ±500 µV ±0.1% of reading 1 -5 5 -5 5 ±1250 µV ±0.1% of reading
PP242 – Current Ranges Shunt index Range Standalone [A] min Range Standalone [A] max Range Thales [A] min Range Thales [A] max Tolerance 0 -40 40 -40 40 ±400 mA ±0.2% of reading 1 -40 40 -40 40 ±40 mA ±0.2% of reading 2 -2 2 -2 2 ±2 mA ±0.2% of reading 3 -200m 200m -200m 200m ±200 µA ±0.2% of reading