The ZENNIUM XC is a compact potentiostat / galvanostat including a frequency response analyser. It offers 32 bit DC resolution and 24 bit AC resolution across 12 different current ranges, which ensure exceptional accurate and precise measurements. The ZENNIUM XC combines compact design with the high-end measurements hardware.

A list of main electrochemical measurements possible with the ZENNIUM XC potentiostat is provided here.
  • DC voltage ranges: ±5 V and ±14 V
  • Current up to ±2 A over 12 current ranges
  • EIS frequency range from 10 µHz to 5 MHz
  • AC amplitude of 0 - 6 V or 0 - 2 A for EIS
  • Switchable floating / grounded mode
Overall bandwidth
DC - 6 MHz
ADC resolution
32 bit
Harmonic reject*
> 60 dB @ ½ full scale
Potentiostat modes
potentiostatic, galvanostatic, pseudo-galvanostatic, OCP, ZRA, FRA
Cell connection
2-, 3-, 4-terminal kelvin
Ground reference
grounded, floating
PC interface
USB 2.0
Dimensions (H x W x D)
160 x 255 x 385 mm3
Weight
8 kg
Accessories
cell cable set, USB cable, power cord, manual, Thales XT software suite, 3 Zahner Analysis licenses
Power supply
100/115/230 VAC, 50/60 Hz
Ambient temperature
+10 °C to +30 °C
Ambient humidity
< 60% without derating
Storage temperature
​-20 °C to +85 °C
*
* Preliminary
Maximum input voltage
±5 V

±14 V
5 V range

14 V range
Voltage resolution
±2.5 nV

±7.5 nV
5 V range

14 V range
Voltage accuracy*
±100 µV ± 5 ppm of reading

±300 µV ± 10 ppm of reading
5 V range

14 V range
DC current resolution
±2 aA (32 Bit)
Main / HiZ Probe
DC current accuracy*
±0.05% of reading ± 0.04% of FS

±0.5% of reading ± 0.4% of FS

±0.5% of reading ± 0.4% of FS

±0.5% of reading ± 125 fA
1 µA ... 100 mA

100 mA ... 2 A

10 nA ... 1 µA

< |1 nA| (HiZ Probe)
Input impedance*
> 10 TΩ || ±5 pF (typical)

> 1000 TΩ || ±1 pF (typical)
Main

HiZ Probe
Input leakage current*
< |200 fA| typ.

< 2 - 5 pA

< |10 fA| typ.
Main



HiZ-Probe
Impedance range*
1 mΩ to 10 GΩ / 2%

100 mΩ to 10 MΩ / 0.2%

100 mΩ to 100 GΩ / 2%

10 µΩ to 1 GΩ / 2%

1 mΩ to 10 MΩ / 0.2%
Main



HiZ

Gal
Common mode rejection*
> 86 dB @ 10 µHz to 100 kHz

> 66 dB @ 100 kHz to 5 MHz
Input channel phase-tracking accuracy
±0.05 ° @ 10 µHz to 100 kHz

±0.125 ° @ 100 kHz to 5 MHz
Equivalent effective input noise*
1 µV rms / 100 fA rms @ 1 mHz to 10 Hz
Main
*
* Preliminary
Controlled voltage
±5 V

±14 V
Resolution
2.5 nV

7.5 nV
5 V range

14 V range
Accuracy*
±200 µV ± 10 ppm of reading

±600 µV ± 10 ppm of reading
5 V range

14 V range
Integral Nonlinearity*
typ. 4 ppm, max. 8 ppm

typ. 12 ppm, max. 24 ppm
5 V range

14 V range
AC amplitude
0 to 2 V, 24 bit resolution

0 to 6 V, 24 bit resolution
5 V range

14 V range
Bandwidth
6 MHz @ 33 Ω load
IR compensation
auto AC impedance technique

range 0 to 10 MΩ

resolution 0.012%
Small signal rise time
150 ns to 200 µs in 5 steps
automatic selection
Slew rate*
15 MV/s
Phase shift*
10° @ 500 kHz
*
* Preliminary
Controlled current
±2 A
Current range
±1.9 nA to ±2 A in 12 steps
Main
Resolution
32 bit, ±0.2 ppb of FS (1ppm @ 1 pA)
Accuracy*
±0.1% of reading ± 0.04 % of FS

±0.4% of reading ± 0.2% of FS
≥ 1 µA to 100 mA

< 1 µA or > 100 mA
*
* Preliminary
Frequency range
10 µHz to 5 MHz
Accuracy*
< 0.0025%
Resolution
0.0025%,

10000 steps/decade
*
* Preliminary
Overall bandwidth
DC - 6 MHz
ADC resolution
32 bit
Harmonic reject*
> 60 dB @ ½ full scale
Potentiostat modes
potentiostatic, galvanostatic, pseudo-galvanostatic, OCP, ZRA, FRA
Cell connection
2-, 3-, 4-terminal kelvin
Ground reference
grounded, floating
PC interface
USB 2.0
Dimensions (H x W x D)
160 x 255 x 385 mm3
Weight
8 kg
Accessories
cell cable set, USB cable, power cord, manual, Thales XT software suite, 3 Zahner Analysis licenses
Power supply
100/115/230 VAC, 50/60 Hz
Ambient temperature
+10 °C to +30 °C
Ambient humidity
< 60% without derating
Storage temperature
​-20 °C to +85 °C
*
* Preliminary
Maximum input voltage
±5 V

±14 V
5 V range

14 V range
Voltage resolution
±2.5 nV

±7.5 nV
5 V range

14 V range
Voltage accuracy*
±100 µV ± 5 ppm of reading

±300 µV ± 10 ppm of reading
5 V range

14 V range
DC current resolution
±2 aA (32 Bit)
Main / HiZ Probe
DC current accuracy*
±0.05% of reading ± 0.04% of FS

±0.5% of reading ± 0.4% of FS

±0.5% of reading ± 0.4% of FS

±0.5% of reading ± 125 fA
1 µA ... 100 mA

100 mA ... 2 A

10 nA ... 1 µA

< |1 nA| (HiZ Probe)
Input impedance*
> 10 TΩ || ±5 pF (typical)

> 1000 TΩ || ±1 pF (typical)
Main

HiZ Probe
Input leakage current*
< |200 fA| typ.

< 2 - 5 pA

< |10 fA| typ.
Main



HiZ-Probe
Impedance range*
1 mΩ to 10 GΩ / 2%

100 mΩ to 10 MΩ / 0.2%

100 mΩ to 100 GΩ / 2%

10 µΩ to 1 GΩ / 2%

1 mΩ to 10 MΩ / 0.2%
Main



HiZ

Gal
Common mode rejection*
> 86 dB @ 10 µHz to 100 kHz

> 66 dB @ 100 kHz to 5 MHz
Input channel phase-tracking accuracy
±0.05 ° @ 10 µHz to 100 kHz

±0.125 ° @ 100 kHz to 5 MHz
Equivalent effective input noise*
1 µV rms / 100 fA rms @ 1 mHz to 10 Hz
Main
*
* Preliminary
Controlled voltage
±5 V

±14 V
Resolution
2.5 nV

7.5 nV
5 V range

14 V range
Accuracy*
±200 µV ± 10 ppm of reading

±600 µV ± 10 ppm of reading
5 V range

14 V range
Integral Nonlinearity*
typ. 4 ppm, max. 8 ppm

typ. 12 ppm, max. 24 ppm
5 V range

14 V range
AC amplitude
0 to 2 V, 24 bit resolution

0 to 6 V, 24 bit resolution
5 V range

14 V range
Bandwidth
6 MHz @ 33 Ω load
IR compensation
auto AC impedance technique

range 0 to 10 MΩ

resolution 0.012%
Small signal rise time
150 ns to 200 µs in 5 steps
automatic selection
Slew rate*
15 MV/s
Phase shift*
10° @ 500 kHz
*
* Preliminary
Controlled current
±2 A
Current range
±1.9 nA to ±2 A in 12 steps
Main
Resolution
32 bit, ±0.2 ppb of FS (1ppm @ 1 pA)
Accuracy*
±0.1% of reading ± 0.04 % of FS

±0.4% of reading ± 0.2% of FS
≥ 1 µA to 100 mA

< 1 µA or > 100 mA
*
* Preliminary
Frequency range
10 µHz to 5 MHz
Accuracy*
< 0.0025%
Resolution
0.0025%,

10000 steps/decade
*
* Preliminary

Downloads

Download the latest information flyer about Zahner ZENNIUM XC potentiostat...

Request a quote

Is this the right product for you? Contact our sales team to get further information about our products and their suitability to your individual applications.

Contact Sales