Electrochemistry

Our Electrochemical Workstations ZENNIUM X / ZENNIUM Pro / ZENNIUM  base on an universal and modular data acquisition system. Together with the THALES software package we offer all standard electrochemical methods at a mouse click.

slider

Photo-Electrochemistry

CIMPS is an universal photo- and spectro-electrochemical workstation for a wide field of applications. You can easily extend the basic CIMPS system with various options for special fields of photo-electrochemical research.

slider

Energy Storage & Conversion

Our power potentiostats, electronic loads and multi-channels acquisition systems provide a smart system for research and investigation on batteries, fuel cells and super caps.

slider

Corrosion

Electrochemical noise has become popular as a new method of corrosion detection and prevention. ZAHNER has developed the CorrElNoise® technique, the first method to obtain both, potential noise and current noise from one electrochemical source.

slider
profile_overview_image

Electrochemical Workstation Zennium E4

Electrochemical Workstation

Compact High Performance

 

The ZENNIUM E4 is a high-performance electrochemical workstation at a surprising affordable price. Besides a potentiostat / galvanostat it bears a frequency response analyzer unit (FRA) for promoting impedance spectroscopy. It provides high-end features like a frequency range up to 4 MHz, a max. current of ±2.0 A, a controlled voltage of ±12 V and an input impedance of 10 TΩ || 5 pF. The ZENNIUM E4 is a perfect workstation for a wide range of electrochemical applications with focus on corrosion, coatings and laminates.

It can be extended by a number of ZAHNER probes.

ZENNIUM E4 was developed using our 35 years of experience in producing high-precision electrochemical workstations and potentiostats / galvanostats of the high-end class. The ZENNIUM E4 is a revolution in the field of compact electrochemical workstations. Since it is equipped with a freqency response analyzer (FRA) unit on-board for the immediate promotion of impedance spectroscopy it can be seen as a unique tool for nearly all kinds of electrochemical measurements during the daily research routine. It provides a wide frequency range from 10 μHz to 4 MHz, an output current up to ±2.0 A and a fast signal processing. Special measurement techniques guarantee an ultra high accuracy and a minimal interference with the test object. The ZENNIUM E4 can be easily switched from grounded to floating operation.

ZENNIUM E4 comes bundled with the outstanding THALES XT software package which offers all standard methods and more at a mouse click. This is why the Zennium E4 can easily be adapted to very different measurement requirements. Furthermore, with the probes available, the ZENNIUM E4 is able to grow with its tasks:

  • Compliance Voltage up to ±120 V (CVB120)
  • Input Impedance up to 1000 TΩ (HiZ Probe)
  • FRA interface to 3rd-party electronic loads or potentiostats (FRA Probe)

General

Overall bandwidth

DC - 5 MHz

ADC resolution

3 ADCs @ 18 bit

Harmonic reject

> 60 dB @ ½ full scale

Potentiostat modes

Potentiostatic, galvanostatic, pseudo-galvanostatic, rest potential, off, ZRA, FRA

Cell connection

2-, 3- ,4-terminal kelvin

Floating / grounded

Switchable

Chassis

Ground

PC interface

USB

Dimensions

255 x 160 x 385 mm

Weight

8 kg

Accessories

Cell cable set, USB cable, power cord, manual

Power supply

230/115/100VAC, 50/60Hz

Ambient temperature, humidity

+15 °C to +35 °C, < 50% without derating

 

Frequency generator & analyzer

Frequency range

10 µHz to 4 MHz

Accuracy

< 0.0025%

Resolution

0.0025%, 10000 steps/decade

 

Output potentiostatic

Controlled voltage

±4 V, ±12 V

 

Resolution

125 µV, 375 µV

 

Accuracy

±500 µV, ±2 mV / ±0.025% of set value

 

Temperature stability

better 20 µV / °C

 

Compliance voltage

±14 V
±120 V

Main
CVB120

AC amplitude

1 mV to 1 V
1 mV to 25 V

Main
CVB120

Bandwidth

2 MHz @ 33 Ω load

 

IR compensation

auto AC impedance technique
range 0 to 10 MΩ
resolution 0.012%



Small signal rise time

250 ns to 200 µs in 5 steps, automatic selection by automatic stability control

 

Slew rate

15 MV/s

 

Phase shift

10° @ 250 kHz

 

 

Output galvanostatic

Full scale current ranges

±100 nA to ±2.0 A in 10 steps, resolution 0.0031 % (16 bit) of range
lowest full scale range ±100 nA, resolution 12.5 pA

Main

 

±1 nA to ±0.5 A in 12 steps, resolution 0.0031 % (16 bit) of range
lowest full scale range ±1 nA, resolution 125 fA

HiZ

Accuracy **

±0.1 % of set value @ >|2 µA| to |100 mA|
±1 % of set value @ |1 nA| to |2 µA| or >|100 mA|
±1 % of set value, ±20 pA @ <|1 nA|

Main

 

±1 % of set value, ±250 fA @ < |1 nA|

HiZ

 

Input

Full scale potential ranges

±1, ±2, ±4 V, ±12 V

 

Potential resolution DC *

0.0008 % / 32 µV (4 V range)
0.0008 % / 100 µV (12 V range)

 

Potential resolution AC *

160 nV

 

Potential accuracy DC **

±0.025 % of reading ±0.25 mV (4 V range)
±0.025 % of reading ±1 mV (12 V range)

 

Offset temperature stability

< 10 µV / °C

 

Full scale current range *

±100 pA to ±2.0 A in 33 ranges,
automatic range selection
±1 pA to ±0.5 A in 35 ranges,
automatic range selection

Main

HiZ

Current accuracy DC **

±0.05 % of reading @ > |2 µA| to |100 mA|
±0.5 % of reading @ < |2 µA| or > |100 mA|
±0.5 % of reading, ±10 pA @ < |1 nA|

Main

 

±0.5 % of reading, ±125 fA @ < |1 nA|

HiZ

Input bias current **

|±1 pA| (typ.) / |±5 pA| (max.)
|±10 fA| (typ.) / |±50 fA| (max.)

Main
HiZ

Current Resolution DC *

2.5 pA
25 fA

Main
HiZ

Current Resolution AC *

16 fA
160 aA

Main
HiZ

Input impedance

10 TΩ || ±5 pF (typ.)
1000 TΩ || ±1 pF (typ.)

Main
HiZ

Impedance range

100 mΩ to 10 MΩ / 0.2%
1 mΩ to 1 GΩ / 2%
30 µΩ to 1 GΩ / 2%
100 mΩ to 100 GΩ / 3%

Main
Pot
Gal
HiZ

Common mode rejection

> 86 dB @ 10 µHz to 100 kHz
> 66 dB @ 100 kHz to 4 MHz

 

Input channel phase-tracking accuracy

±0.1 ° @ 10 µHz to 100 kHz
±0.25 ° @ 100 kHz to 4 MHz

 

Equivalent effective input noise

2 µV rms / 200 fA rms @ 1 mHz to 10 Hz
20 µV rms / 30 fA rms @ 1 mHz to 10 Hz

Main
HiZ

* Garanteed by design. For details refer to application note "how to read specifications".

** In the first 6 months after factory calibration, after 20 min. of warm-up.

All ZENNIUM Electrochemical Workstations specifications PDF data sheet for download.

Zennium E4 Flyer

Download the latest information flyer about ZAHNER Zennium E4 potentiostat...