ZAHNER ZENNIUM X
Electrochemical Workstation
The ultimate high performance
Since 1976, ZAHNER-elektrik is producing Electrochemical Workstations at the edge of technology. The systems are highly praised for their exceptional accuracy, flexibility and reliability. Now, the ZAHNER R&D-team launched a new generation of Electrochemical Workstations including the ZENNIUM X defining the Goldstandard in electrochemical instrumentation.
Going through the technical specifications shown below you will find outstanding features and data like
- 32 bit converter resolution for exceptional DC accuracy
- Electrochemical Impedance Spectroscopy (EIS) frequency range from 10 µHz to 12 MHz
- HDP technology with24 bit AC resolution for highest dynamic accuracy
- ADF technology for extremely fast CV up to 10 kV/s with continuous slopes for outstanding CV performance
- Compliance voltage up to ±32 V
- Software switchable compliance voltage for minimum noise, optimum sample safety and energy efficieny
- DC voltage ranges ±5 V and ±15 V software controlled perfectly matching the requirements of Lithium battery research
- AC amplitude 0 - 2 V and 0 - 6 V with 18 bit resolution
- Modular concept for high flexibility and expandability
- Floating / grounded mode switchable for measuring on a wide range of objects
The ZENNIUM X come with the new THALES XT Software Suite providing features such as highly flexible Windows®-conform online visualization and intuitive offline data analysis.
The ZENNIUM X is a modular electrochemical workstation including a potentiostat / galvanostat and a frequency response analyzer (FRA) unit with a frequency range up to 12 MHz at ±4 A and at a compliance voltage of up to ±32 V. With an input impedance of more than 10 TΩ || 5 pF the workstation is suitable for a wide range of electrochemical tasks.
The electrochemical measurement capabilites of the ZENNIUM X workstation can be extented by adding up to ten additional plug-in cards such as EPC42, TEMP/U or PAD4. For example, the EPC42 allows the ZENNIUM X to control up to four additional external devices such as power potentiostats of the PP/XPot-series (boosters) or electrical loads of the EL-series. Thus, our ZENNIUM X electrochemical workstation is able to provide even more advanced measurement methods and allows us to satisfy any of your electrochemical research requirements.
General |
|
Overall bandwidth |
DC - 15 MHz |
ADC resolution |
32 bit |
Harmonic reject* |
> 60 dB @ ½ full scale |
Potentiostat modes |
potentiostatic, galvanostatic, pseudo-galvanostatic, rest potential, off |
Cell connection |
2-, 3-, 4-terminal kelvin |
Ground reference |
grounded, floating |
Extension slots |
10 |
PC interface |
USB 2.0 |
Dimensions |
470 x 160 x 376 mm |
Weight |
13.2 kg |
Accessories |
cell cable set, USB cable, power cord, manual, Testbox, THALES XT Software Suite |
Power supply |
100/115/230VAC, 50/60Hz |
Ambient temperature |
+10 °C to +30 °C |
Ambient humidity |
< 60% without derating |
Frequency generator & analyzer |
|
Frequency range |
10 µHz to 12 MHz |
Accuracy* |
< 0.0025% |
Resolution |
0.0025%, 10000 steps/decade |
Output potentiostatic |
||
Controlled voltage |
±5 V |
|
Resolution |
2.5 nV |
5 V range |
Accuracy* |
±50 µV ± 5 ppm of reading |
5 V range |
Integral Nonlinearity* |
typ. 1 ppm, max. 2 ppm |
5 V range |
Compliance voltage |
±16 V |
Low compliance voltage mode |
AC amplitude |
0 to 2 V, 18 bit resolution |
5 V range |
Bandwidth |
15 MHz @ 33 Ω load |
|
IR compensation |
auto AC impedance technique |
|
Small signal rise time |
150 ns to 200 µs in 5 steps |
automatic selection |
Slew rate* |
15 MV/s |
|
Phase shift* |
10° @ 500 kHz |
|
Output galvanostatic |
||
Controlled current |
±4 A |
|
Current range |
±1 nA to ±4 A in 12 steps |
Main |
Resolution |
32 bit, ±0.2 ppb of FS |
|
Accuracy* |
0.01 % of FS ± 0.025 % of reading |
≥ 1 µA to 100 mA |
Input |
||
Maximum input voltage |
±5.5 V |
5 V range |
Voltage resolution |
2.5 nV |
5 V range |
Voltage accuracy* |
±20 µV ± 2 ppm of reading |
5 V range |
Current resolution |
±5 aA (32 Bit) |
Main / HiZ Probe |
Current accuracy* |
± 0.01% of FS |
1 µA ... 100 mA |
Input impedance* |
> 10 TΩ || ±5 pF (typical) |
Main |
Input leakage current* |
< |± 200 fA| typ. |
|
Impedance range* |
1 mΩ to 10 GΩ / 1% |
Main |
Common mode rejection* |
> 86 dB @ 10 µHz to 100 kHz |
|
Input channel phase-tracking accuracy* |
±0.05 ° @ 10 µHz to 100 kHz |
|
Equivalent effective input noise* |
1 µV rms / 100 fA rms @ 1 mHz to 10 Hz |
Main HiZ |
*Preliminary
All ZENNIUM Electrochemical Workstations specifications PDF data sheet for download.