Electrochemistry

Our Electrochemical Workstations ZENNIUM X / ZENNIUM Pro / ZENNIUM  base on an universal and modular data acquisition system. Together with the THALES software package we offer all standard electrochemical methods at a mouse click.

slider

Photo-Electrochemistry

CIMPS is an universal photo- and spectro-electrochemical workstation for a wide field of applications. You can easily extend the basic CIMPS system with various options for special fields of photo-electrochemical research.

slider

Energy Storage & Conversion

Our power potentiostats, electronic loads and multi-channels acquisition systems provide a smart system for research and investigation on batteries, fuel cells and super caps.

slider

Corrosion

Electrochemical noise has become popular as a new method of corrosion detection and prevention. ZAHNER has developed the CorrElNoise® technique, the first method to obtain both, potential noise and current noise from one electrochemical source.

slider
profile_overview_image

fF Probe

The femto-Farad Probe can measure very low capacitance with a resolution of 0.1 fF. Due to the transimpedance princible the fF Probe is designated for the application on two-pole objects.

The femto-Farad Probe works as a front-end to the IM6/Zennium-potentiostat. Apart from its limited current capability, all basic functionalities of the Thales software are supported. In particular impedance spectroscopy can be applied. Due to the fact, that the primary measurement magnitude is the complex impedance, besides the sample capacity, resistive and DC contributions can be determined as well.

femto-Farad Probe

 

Frequency range

10 µHz to 1 MHz

Current autoranging,
manually selectable current ranges

0 - ±40 nA,
±40 nA - ±400 nA,
±400 nA - ±4 µA,
±4 µA - ±40 µA

Voltage range

±4 V

Resolution of any range

18 bit

Capacitance offset

±1 fF *)

Capacitance resolution

±0.1 fF *)

Capacitance accuracy

±0.25% of reading ±2 fF *)

 *) current range ±40 nA, AC amplitude ≥ 100 mV, zero DC current

Download the latest manual about ZAHNER femto-Farad Probe...