IM7 - Current ranges

The IM7 potentiostats offer 12 different current ranges for accurately measuring currents across different orders of magnitude. Lists of different current ranges of IM7 potentiostats are provided below.
 
Current ranges of IM7x
 
Range # Shunt [Ω] Max. current  [A]
0 10 m ± 4.0
1 100 m ± 4.0
2 1 ± 600 m
3 10 ± 300 m
4 100 ± 30 m
5 1 k ± 3 m
6 10 k ± 300 µ
7 100 k ± 30 µ
8 1 M ± 3 µ
9 10 M ± 300 n
10 100 M ± 30 n
11 1 G ± 3 n
Current ranges of IM7
 
Range # Shunt [Ω] Max. current  [A]
0 10 m ± 3.0
1 100 m ± 3.0
2 1 ± 600 m
3 10 ± 300 m
4 100 ± 30 m
5 1 k ± 3 m
6 10 k ± 300 µ
7 100 k ± 30 µ
8 1 M ± 3 µ
9 10 M ± 300 n
10 100 M ± 30 n
11 1 G ± 3 n
Current ranges of IM7c
 
Range # Shunt [Ω] Max. current  [A]
0 10 m ± 2.0
1 100 m ± 2.0
2 1 ± 600 m
3 10 ± 300 m
4 100 ± 30 m
5 1 k ± 3 m
6 10 k ± 300 µ
7 100 k ± 30 µ
8 1 M ± 3 µ
9 10 M ± 300 n
10 100 M ± 30 n
11 1 G ± 3 n

32-bit DC resolution: A 32-bit DC resolution in a current range indicates that in the current range, 232 data points can be measured between the max. current limits.
 
For IM7x, IM7 and IM7c potentiostats, the first two current ranges contain the same max. current ranges respectively but two different shunt resistances. A lower shunt resistance (10 mΩ) leads to a lower voltage drop across the shunt resistance. With this, most of the compliance voltage of a potentiostat can be used to drive the test object. With higher shunt resistance (100 mΩ), the voltage drop across the shunt resistance is high. This limits the use of full scale of the compliance voltage but provides higher current resolution. The IM7 potentiostats choose the appropriate shunt resistance itself during the measurement to provide accurate and well-resolved results.

Request a quote

Is this the right product for you? Contact our sales team to get further information about our products and their suitability to your individual applications.

Contact Sales